Ram Meter Inc. B750A50 - 750 Amp 50 DCmV DC Current Shunt

Call for Price
SKU
B750A50

Ram Meter Inc. DC (direct current) instrument shunts are used to extend the range of ammeters whenever the current to be measured is too large to be passed through the instrument - usually currents over 50 amperes. Therefore the shunt is a diverter which is used to "shunt" the majority of the current around the indicating instrument. The shunt is essentially a specially designed resistor which delivers a millivolt output (to a millivolt meter or instrument) in proportion to the current flowing through the resistor.

Features:

  • Available in ranges from 0.5A to 3500A+ (higher ranges available by special request)
  • 0-50 and 0-100 DCmV Standard Outputs
  • Portable, Bus-Bar and Commercial/Military Lightweight styles available
  • Portable type shunts come on insulating bases which can be fastened to any flat surface
  • Also available WITHOUT a base by special request

Adjusted Accuracy
Ram Meter Inc. shunts are adjusted to within +/-0.25% nominal value with an applied current that will not raise the temperature of the terminal blocks more than 50°F (10°C) with an ambient room temperature of 77°F (25°C) +/-2°.

Rated Accuracy
Commercial shunts up to 1,000 amps have an accuracy of +/-0.33%
Commercial shunts 1,000 amps and above have an accuracy of +/-0.5%
Commercial lightweight shunts have an accuracy of +/-0.6%
Military lightweight shunts have an accuracy of +/-0.6%

This allows for self-heating variation at the higher current values and variance in connections.

What is the difference between Adjusted & Rated Accuracy?
The Adjusted Accuracy is when the shunt is being calibrated (adjusted) for the first time after it is manufactured. This is done at room temperature (77°F / 25°C) with a current low enough to not raise the temperature of the shunt. So if the shunt is used at 77°F / 25°C it will be this accurate.
The Rated Accuracy is how accurate the shunt will be with up to 2/3 of the rated current running through it continuously. In this case the shunt will heat up and the resistance can change as it heats up. Therefore, the accuracy will change to be within the rated accuracy.

CUSTOM SHUNT DESIGN

We welcome the opportunity to work with you on your special design applications where other current, millivolt, size, accuracy or environmental conditions preclude the use of our standard catalog items. If this is the case, please fill out our Ram Meter Inc. Special Shunt Requirement Form or CONTACT US and one of our highly trained sales representatives will help you find a solution.

For all other shunt-related questions, check out our detailed RAM METER INC. SHUNT DISCUSSION.

More Information
Input 750 DCA
Features Brass End Block
Manufacturer Ram Meter Inc.
Output 50 DCmV
Type Current Shunts